dc.identifier.citationHeras-Mulero H, Moreno-Montanes J, Sadaba Echarri LM, Mendiluce Martin L. Comparison of dynamic contour tonometry (Pascal) with pneumotonometry and Goldmann tonometry. Arch Soc Esp Oftalmol 2007 Jun;82(6):337-341es_ES
dc.identifier.issn0365-6691-
dc.identifier.urihttps://hdl.handle.net/10171/23331-
dc.description.abstractTo compare the intraocular pressure measurements as defined by the Pascal tonometer, the Goldmann tonometer and the pneumotonometer. METHODS: This was an observational clinical study, which included two hundred and five randomly selected subjects recruited from the Ophthalmology Department. The intraocular pressure measurements were performed with each tonometry technique in a randomized order. RESULTS: The Pascal's intraocular pressure measurement was significantly higher than that measured by the other two tonometers (p<0.05). The quality data of Pascal was: optimum in 27.3% (56 of 205 patients), acceptable in 42% (86 of 205 patients) and unacceptable in 23.4% (48 of 205 patients). In 7.3% (15 of 205 patients) it was impossible to obtain any measurement using Pascal. A weak correlation coefficient between the Pascal and the Goldmann, and between Pascal and the pneumotonometer was found. The Bland-Altman method of measurement using these tonometers showed a high degree of discordance. CONCLUSION: As reported by others authors, the Pascal's intraocular pressure measurement is higher than that of the Goldmann tonometer. The measurement differs from 0.7 to 4.4 mmHg. In corneas with pathology, it is very difficult or even unacceptable to measure the intraocular pressure using the Pascal tonometer.es_ES
dc.language.isoenges_ES
dc.publisherElsevier Españaes_ES
dc.rightsinfo:eu-repo/semantics/openAccesses_ES
dc.subjectTonometry, Ocular/instrumentationes_ES
dc.subjectPpneumotonometryes_ES
dc.titleComparison of dynamic contour tonometry (Pascal®) with pneumotonometry and Goldmann tonometryes_ES
dc.title.alternativeComparación del tonómetro Pascal® con el neumotonómetro y el tonómetro Goldmannes_ES
dc.typeinfo:eu-repo/semantics/articlees_ES
dc.relation.publisherversionhttp://www.oftalmo.com/seo/archivos/maquetas/A/7CC3A6F7-11D6-7E88-AB90-000065858ACA/articulo.pdfes_ES
dc.type.driverinfo:eu-repo/semantics/articlees_ES

Files in This Item:
Thumbnail
File
articulo-1.pdf
Description
Size
195.19 kB
Format
Adobe PDF


Statistics and impact
0 citas en
0 citas en

Items in Dadun are protected by copyright, with all rights reserved, unless otherwise indicated.