Influence of the test-chamber shape on the performance of conductometric gas sensors
Keywords: 
Computational fluid dynamics (CFD)
Conductometric gas sensor
Gas flow
Chamber-position influence
Issue Date: 
2022
Publisher: 
Elsevier
ISSN: 
0925-4005
Note: 
This is an open access article under the CC BY-NC-ND license
Citation: 
Parellada-Monreal, L. (Laura); Garcia-Mandayo, G. (Gemma); Martín-Mayor, A. (Alain); et al. "Influence of the test-chamber shape on the performance of conductometric gas sensors". Sensors and Actuators B: Chemical. 361, 2022, 131694
Abstract
In this article, CFD simulations results are presented as a key tool to the comprehension of the target gas con- centration evolution in a test chamber, at different working conditions. The simulation results are compared with the experimental data, which shows a qualitative good correlation with the evolution of the concentration gradient detected. The experiments were carried out using an aluminum gas test chamber, where a WO3 based conductometric sensor is introduced. The results demonstrate how the response time is dependent on the sensor working conditions. Analyzing the CFD and experimental results, some assumptions for this behavior are proposed. The WO3 sensor needs a Pt heating element, which is heated up to 300 ◦C. As the response is highly temperature-dependent, the temperature distribution on the sensor surface was measured by an IR thermo- graphic camera. The simulation results show that the temperature distribution matches with those obtained experimentally. To validate the model, a mesh and time step convergence study was also implemented

Files in This Item:
Thumbnail
File
1-s2.0-S0925400522003367-main.pdf
Description
Size
6.31 MB
Format
Adobe PDF


Statistics and impact
0 citas en
0 citas en

Items in Dadun are protected by copyright, with all rights reserved, unless otherwise indicated.