Full metadata record
DC FieldValueLanguage
dc.creatorParellada-Monreal, L. (Laura)-
dc.creatorGarcia-Mandayo, G. (Gemma)-
dc.creatorMartín-Mayor, A. (Alain)-
dc.creatorSouto-Canteli, I. (Íñigo)-
dc.creatorBou-Ali, M.M. (M. Mounir)-
dc.date.accessioned2022-08-01T10:47:25Z-
dc.date.available2022-08-01T10:47:25Z-
dc.date.issued2022-
dc.identifier.citationParellada-Monreal, L. (Laura); Garcia-Mandayo, G. (Gemma); Martín-Mayor, A. (Alain); et al. "Influence of the test-chamber shape on the performance of conductometric gas sensors". Sensors and Actuators B: Chemical. 361, 2022, 131694es
dc.identifier.issn0925-4005-
dc.identifier.urihttps://hdl.handle.net/10171/63850-
dc.description.abstractIn this article, CFD simulations results are presented as a key tool to the comprehension of the target gas con- centration evolution in a test chamber, at different working conditions. The simulation results are compared with the experimental data, which shows a qualitative good correlation with the evolution of the concentration gradient detected. The experiments were carried out using an aluminum gas test chamber, where a WO3 based conductometric sensor is introduced. The results demonstrate how the response time is dependent on the sensor working conditions. Analyzing the CFD and experimental results, some assumptions for this behavior are proposed. The WO3 sensor needs a Pt heating element, which is heated up to 300 ◦C. As the response is highly temperature-dependent, the temperature distribution on the sensor surface was measured by an IR thermo- graphic camera. The simulation results show that the temperature distribution matches with those obtained experimentally. To validate the model, a mesh and time step convergence study was also implementedes_ES
dc.description.sponsorshipMondragon Goi Eskola Politeknikoa (MGEP) and Ceit thank Elkartek Project KK-2021/00082 (micro4IloT) and IT1505-22 (Research Group Program) of the Basque Country Government for funding the research activities in this paperes_ES
dc.language.isoenges_ES
dc.publisherElsevieres_ES
dc.rightsinfo:eu-repo/semantics/openAccesses_ES
dc.subjectComputational fluid dynamics (CFD)es_ES
dc.subjectConductometric gas sensores_ES
dc.subjectGas flowes_ES
dc.subjectChamber-position influencees_ES
dc.titleInfluence of the test-chamber shape on the performance of conductometric gas sensorses_ES
dc.typeinfo:eu-repo/semantics/articlees_ES
dc.description.noteThis is an open access article under the CC BY-NC-ND licensees_ES
dc.identifier.doi10.1016/j.snb.2022.131694-
dadun.citation.publicationNameSensors and Actuators B: Chemicales_ES
dadun.citation.startingPage131694es_ES