Long-term voice monitoring., Wearable system, Phonatory disorders

Abstract

The objective quantification of voice acoustic parameters is used for the diagnosis, monitoring, and treatment of voice disorders. Such assessments are carried out with specialised equipment within the doctor's office. The controlled conditions employed are usually not those of the real environment of the patient. The results, although very informative, are specific to those measurement conditions and to the time when they were performed. A wearable voice monitoring system, based on an accelerometer to ensure the message, can overcome these limitations. We present a miniaturised, low-power, and low-cost wearable system to estimate and record voice fundamental frequency (F0), intensity and phonation time for long intervals in the everyday environment of the patient. It was tested on two subjects for up to two weeks of recording time. It was possible to identify distinct periods in vocal activity, such as normal, professional, demanding or hyperfunctional. It provided information on the workload that the vocal cords needed to cope with over time and when and to what extent that workload was concentrated. The proposed voice dosimetry system enables the extraction and recording of voice parameters for long periods of time in the everyday environment of the patient, allowing the objectification of vocal risk situations and personalised treatment and monitoring.