Catastrophic wear in a metal-on-ceramic total hip arthroplasty
Keywords: 
Arthroplasty, Replacement, Hip/instrumentation
Hip Prosthesis/adverse effects
Prosthesis Design
Issue Date: 
2007
Publisher: 
Elsevier
ISSN: 
0883-5403
Citation: 
Valentí JR, Del Río J, Amillo S. Catastrophic wear in a metal-on-ceramic total hip arthroplasty. J Arthroplasty. 2007 Sep;22(6):920-2.
Abstract
A 51-year-old woman came to our clinic 6 months after a right total hip arthroplasty. She had noticed a slowly growing mass in the proximal thigh and referred progressive pain in the right groin. Plain radiography revealed premature acetabular cup aseptic loosening, and in the computed tomography study, a 14-cm-diameter mass was observed. Revision surgery was performed, showing a metal-on-ceramic bearing surface. The histologic analysis of surrounding tissues was reported as massive metallosis. Although occasionally chosen for primary or revision hip arthroplasty, there is little information available about the in vivo wear behavior of this combination. This important fact should be taken into account before considering such a surface alternative

Files in This Item:
Thumbnail
File
J Arthroplasty.2007.920.pdf
Description
Size
654.14 kB
Format
Adobe PDF


Statistics and impact
0 citas en
0 citas en

Items in Dadun are protected by copyright, with all rights reserved, unless otherwise indicated.