Effective dose estimation for oncological and neurological PET/CT procedures
Keywords: 
PET/CT
Effective dose
Radiopharmaceutical
Whole body
Issue Date: 
2017
Publisher: 
Springer
ISSN: 
2191-219X
Note: 
This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
Citation: 
Marti-Climent, J.M. (Josep María); Prieto, E. (Elena); Moran, V. (Verónica); et al. "Effective dose estimation for oncological and neurological PET/CT procedures". EJNMMI Research. 7 (1), 2017, 37
Abstract
Background: The aim of this study was to retrospectively evaluate the patient effective dose (ED) for different PET/ CT procedures performed with a variety of PET radiopharmaceutical compounds. PET/CT studies of 210 patients were reviewed including Torso (n = 123), Whole body (WB) (n = 36), Head and Neck Tumor (HNT) (n = 10), and Brain (n = 41) protocols with 18FDG (n = 170), 11C-CHOL (n = 10), 18FDOPA (n = 10), 11C-MET (n = 10), and 18F-florbetapir (n = 10). ED was calculated using conversion factors applied to the radiotracer activity and to the CT dose-length product. Results: Total ED (mean ± SD) for Torso-11C-CHOL, Torso-18FDG, WB-18FDG, and HNT-18FDG protocols were 13.5 ± 2.2, 16.5 ± 4.5, 20.0 ± 5.6, and 15.4 ± 2.8 mSv, respectively, where CT represented 77, 62, 69, and 63% of the protocol ED, respectively. For 18FDG, 18FDOPA, 11C-MET, and 18F-florbetapir brain PET/CT studies, ED values (mean ± SD) were 6.4 ± 0.6, 4.6 ± 0.4, 5.2 ± 0.5, and 9.1 ± 0.4 mSv, respectively, and the corresponding CT contributions were 11, 14, 23, and 26%, respectively. In 18FDG PET/CT, variations in scan length and arm position produced significant differences in CT ED (p < 0.01). For dual-time-point imaging, the CT ED (mean ± SD) for the delayed scan was 3.8 ± 1.5 mSv. Conclusions: The mean ED for body and brain PET/CT protocols with different radiopharmaceuticals ranged between 4.6 and 20.0 mSv. The major contributor to total ED for body protocols is CT, whereas for brain studies, it is the PET radiopharmaceutical.

Files in This Item:
Thumbnail
File
s13550-017-0272-5.pdf
Description
Size
370.66 kB
Format
Adobe PDF


Statistics and impact
0 citas en
0 citas en

Items in Dadun are protected by copyright, with all rights reserved, unless otherwise indicated.