Application of the Monte Carlo efficiency transfer method to an HPGe detector with the purpose of environmental samples measurement
Palabras clave : 
HPGe detector
Monte Carlo
Efficiency transfer
Low-background gamma spectrometry
Fecha de publicación : 
2015
Editorial : 
Elsevier
ISSN : 
1872-9800
Cita: 
Morera-Gómez, Y. (Yasser); Cartas-Aguila, H.A. (Héctor A.); Alonso-Hernández, C. M. (Carlos Manuel); et al. "Application of the Monte Carlo efficiency transfer method to an HPGe detector with the purpose of environmental samples measurement". Applied Radiation and Isotopes. 97, 2015, 59 - 62
Resumen
Monte Carlo efficiency transfer method was used to determine the full energy peak efficiency of a coaxial n-type HPGe detector. The efficiencies calibration curves for three Certificate Reference Materials were determined by efficiency transfer using a 152Eu reference source. The efficiency values obtained after efficiency transfer were used to calculate the activity concentration of the radionuclides detected in the three materials, which were measured in a low-background gamma spectrometry system. Reported and calculated activity concentration show a good agreement with mean deviations of 5%, which is satisfactory for environmental samples measurement.